Getting Started Manual LabMaster 9 Zi-A Oscilloscopes
LabMaster 9 Zi-A 2 922157-00 Rev A Safety Instructions This section contains instructions that must be observed to keep the instrument operating in a
LabMaster 9 Zi-A 92 922157-00 Rev A Shortcut Buttons Several dialogs contain common functions accessible from a row of buttons that can save you a ste
Getting Started Manual 922157-00 Rev A 93 Label - Enables you to attach identifying labels to your waveforms. The labels are preserved when the wav
LabMaster 9 Zi-A 94 922157-00 Rev A Turning on Channels and Traces Turn on a channel, math, or memory trace using the Front Panel channel buttons, or
Getting Started Manual 922157-00 Rev A 95 Timebase Timebase Overview You can access Timebase settings using the front panel Horizontal controls, the T
LabMaster 9 Zi-A 96 922157-00 Rev A Mark Don't Show Me This Message Again to avoid subsequent warnings. The 2.92 mm connector saver pop-up. The
Getting Started Manual 922157-00 Rev A 97 When selected, the respective channels are combined into one channel at higher bandwidth. The combined chann
LabMaster 9 Zi-A 98 922157-00 Rev A When DBI is selected, trace descriptor labels correspond with your choices. So, a 30 or 36 GHz example where you d
Getting Started Manual 922157-00 Rev A 99 SLAVE ACQUISITION MODULE DBI SETTINGS When you use DBI and make correct connections and channel selections f
LabMaster 9 Zi-A 100 922157-00 Rev A Sampling Modes Sampling Modes Overview Depending on your timebase, you can choose Single-shot Sampling Mode (belo
Getting Started Manual 922157-00 Rev A 101 You can sample the waveform in a range starting well before the trigger event up to the moment the event oc
Getting Started Manual 922157-00 Rev A 3 Precautions Use proper power cord. Use only the power cord shipped with this instrument and certified for the
LabMaster 9 Zi-A 102 922157-00 Rev A You can combine sequence mode with an advanced trigger to isolate a rare event, capture all instances over hou
Getting Started Manual 922157-00 Rev A 103 Sequence Display Modes The instrument gives you a choice of five ways to display your segments: Adjacent
LabMaster 9 Zi-A 104 922157-00 Rev A Sequence Mode Setup When setting up Sequence Mode, you define the number of fixed-size segments acquired in singl
Getting Started Manual 922157-00 Rev A 105 4. Touch the Enable Timeout checkbox. 5. Touch inside the Timeout data entry control and provide a timeou
LabMaster 9 Zi-A 106 922157-00 Rev A Zooming Segments in Sequence Mode You can zoom individual segments easily using the QUICKZOOM front panel button.
Getting Started Manual 922157-00 Rev A 107 zooming. Now with each touch of the zoom control buttons, the degree of magnification changes by small incr
LabMaster 9 Zi-A 108 922157-00 Rev A Viewing Time Stamps You can view time stamps for each segment. View Segment Time Stamps 1. Touch Timebase → Acq
Getting Started Manual 922157-00 Rev A 109 time between ADC sampling instants and the event trigger provides the necessary variation. The instrument r
LabMaster 9 Zi-A 110 922157-00 Rev A Roll Mode Roll mode displays, in real time, incoming points in single-shot acquisitions that have a sufficiently
Getting Started Manual 922157-00 Rev A 111 Vertical You can access Vertical settings using the front panel Vertical controls, by selecting Vertical→ C
LabMaster 9 Zi-A 4 922157-00 Rev A Cooling The instrument relies on forced air cooling with internal fans and vents. Take care to avoid restricting th
LabMaster 9 Zi-A 112 922157-00 Rev A Vertical Offset - For each channel, you can select between zero vertical offset or to set the offset to a spec
Getting Started Manual 922157-00 Rev A 113 Setting Up Channels 1. Touch Vertical → Channelx Setup... from the menu bar. 2. Touch inside the Trace On
LabMaster 9 Zi-A 114 922157-00 Rev A Shortcut Buttons Shortcut buttons provide quick access to various functions for your respective channel. The foll
Getting Started Manual 922157-00 Rev A 115 Pre-Processing Controls Pre-Processing is meant as before Math processing. Pre-Processing controls provide
LabMaster 9 Zi-A 116 922157-00 Rev A found on the cable's data sheet) in your test setup and analyze your signal with the effects of the cables r
Getting Started Manual 922157-00 Rev A 117 Response Optimization Modes Frequency response and group delay of the oscilloscope contributes to the pulse
LabMaster 9 Zi-A 118 922157-00 Rev A Eye Diagram HBR (High Bit Rate) - same as Eye Diagram, but optimized for high bit rate signals (bit rate >
Getting Started Manual 922157-00 Rev A 119 LabMaster 9 Zi-A Channel Setup You can set up to 20 channels on the Channel Setup dialog. You can make adju
LabMaster 9 Zi-A 120 922157-00 Rev A LabMaster De-Skew Calibration The ChannelSync™ architecture in LabMaster modular oscilloscope systems utilizes a
Getting Started Manual 922157-00 Rev A 121 There are two different deskew calibrations that can be performed. Both permit quick and easy deskew of all
Getting Started Manual 922157-00 Rev A 5 Power AC Power Source 100 to 240 VAC (+/-10%) at 50/60 Hz (+/-10%) Manual voltage selection is not required
LabMaster 9 Zi-A 122 922157-00 Rev A 2. Touch the SystemSkew Cal button. The System Channel/Trigger Skew Calibration Wizard opens. 3. Connect the
Getting Started Manual 922157-00 Rev A 123 4. When the previous operation is complete, you will be prompted to use the same cable as used to connect
LabMaster 9 Zi-A 124 922157-00 Rev A If you perform ChannelSync Cal on multiple gain settings, perform it first on the 100 mV/div setting, then per
Getting Started Manual 922157-00 Rev A 125 1. From the menu bar, choose Vertical > Channel <#> Setup. Touch the ChannelSync Cal tab to the r
LabMaster 9 Zi-A 126 922157-00 Rev A As with System Skew calibration, you will see the Fast Edge output signal appear on the display and delay (skew)
Getting Started Manual 922157-00 Rev A 127 Trigger NOTE: Channels 1-4 have complete triggering capability for Edge, SMART and other triggers. Channel
LabMaster 9 Zi-A 128 922157-00 Rev A Trigger Types There are a set of standard Trigger Types. These Triggers are basic waveform features or conditions
Getting Started Manual 922157-00 Rev A 129 Pattern (Logic) Logic trigger enables triggering on a logical combination (pattern) of five inputs: CH1, C
LabMaster 9 Zi-A 130 922157-00 Rev A GLITCH Glitch trigger is a simpler form of Width trigger. Use Glitch trigger when you want to define a fixed pul
Getting Started Manual 922157-00 Rev A 131 Measurement This trigger type allows you to leverage parameter measurements as waveform trigger conditions
LabMaster 9 Zi-A 6 922157-00 Rev A Power and Ground Connections The 9xxMZi-A Master Acquisition Module is provided with a 15A/250V 14AWG rated ground
LabMaster 9 Zi-A 132 922157-00 Rev A QUALIFIED A-B In single trigger mode, Qualify A-B arms the oscilloscope on the A event, and then triggers on the
Getting Started Manual 922157-00 Rev A 133 Trigger Settings Horizontal and Vertical adjustments are typically made for all trigger types using either
LabMaster 9 Zi-A 134 922157-00 Rev A Trigger Setup Based on your Trigger Type selection, sections of the main Trigger dialog and additional dialogs al
Getting Started Manual 922157-00 Rev A 135 3. If available, touch inside the Coupling control and select a coupling mode. Coupling refers to the typ
LabMaster 9 Zi-A 136 922157-00 Rev A Edge trigger works on the selected edge at the chosen level. The slope (positive or negative) is specified on th
Getting Started Manual 922157-00 Rev A 137 Note that the trigger level is specified in volts and normally remains unchanged when the vertical gain or
LabMaster 9 Zi-A 138 922157-00 Rev A Width Condition is Width triggers (and other triggers equipped with conditional logic) have conditional logic set
Getting Started Manual 922157-00 Rev A 139 Use holdoff to obtain a stable trigger for repetitive, composite waveforms. For example, if the number or d
LabMaster 9 Zi-A 140 922157-00 Rev A Edge Trigger with Holdoff by Events (in this example, two events). The bold edges on the trigger source indicate
Getting Started Manual 922157-00 Rev A 141 TriggerScan TriggerScan is a debugging tool (available for any trigger type) that helps you quickly find ra
Getting Started Manual 922157-00 Rev A 7 Always use the Standby button or the File > Shutdown menu option to execute a proper shut down process and
LabMaster 9 Zi-A 142 922157-00 Rev A Training TriggerScan The TriggerScan Trainer inspects a currently acquired waveform and automatically builds a li
Getting Started Manual 922157-00 Rev A 143 Starting TriggerScan After you have run the Trainer, the Trigger List displays a list of smart trigger setu
LabMaster 9 Zi-A 144 922157-00 Rev A PLEASE NOTE THE FOLLOWING: You can tune the dwell time that the oscilloscope waits before loading the next tri
Getting Started Manual 922157-00 Rev A 145 Viewing Waveforms Display Oscilloscope display grids present the trace information in a voltage vs. time fo
LabMaster 9 Zi-A 146 922157-00 Rev A Display Setup To adjust the appearance of the oscilloscope display: 1. Touch Display → Display Setup. 2. Touch
Getting Started Manual 922157-00 Rev A 147 Sequence Mode Display Set up a Sequence Mode display by first selecting a Sequence trigger mode in the Time
LabMaster 9 Zi-A 148 922157-00 Rev A Dual Display Grid Selection Make a dual grid display selection for showing on an external display (once properly
Getting Started Manual 922157-00 Rev A 149 accumulate on-screen points from many acquisitions to see your signal change over time. The instrument pers
LabMaster 9 Zi-A 150 922157-00 Rev A 2. Touch one of the persistence mode buttons next to an input channel. To turn off persistence for an individua
Getting Started Manual 922157-00 Rev A 151 WaveStream mode operates at up to 80 GS/s with an update rate of up to several thousand waveforms/second fo
LabMaster 9 Zi-A 8 922157-00 Rev A LabMaster Overview LabMaster is a unique modular oscilloscope solution that allows a configuration of more channels
LabMaster 9 Zi-A 152 922157-00 Rev A Pushing the Intensity knob changes the Persistence status in the following ways. OFF - When the button hasn&a
Getting Started Manual 922157-00 Rev A 153 Zooming Waveforms Zooming Waveforms Overview You can magnify a selected region of a waveform using the Zoom
LabMaster 9 Zi-A 154 922157-00 Rev A When you zoom a waveform, a preview of the zoomed area is shown on the Preview section of the Zx dialog. Zooming
Getting Started Manual 922157-00 Rev A 155 by a small increment. OR If you want to zoom in or out in large standard increments with each touch of the
LabMaster 9 Zi-A 156 922157-00 Rev A Touch-and-Drag Zooming 1. Touch and drag a rectangle around any part of an input channel waveform, math trace, o
Getting Started Manual 922157-00 Rev A 157 Measuring with Cursors Cursors are important tools that aid you in measuring signal values. Cursors are mar
LabMaster 9 Zi-A 158 922157-00 Rev A Cursor Setup After using one of the above methods to access cursors, the Cursors dialog is displayed. Cursor Typ
Getting Started Manual 922157-00 Rev A 159 cursors then move together and maintaining their relative distance from their originally specified position
LabMaster 9 Zi-A 160 922157-00 Rev A Measurement Parameters Parameters are measurement tools that determine a wide range of waveform properties. Use t
Getting Started Manual 922157-00 Rev A 161 Parameter Setup 1. Touch the parameter tab (Px) of an unused location or one that you want to change. 2.
Getting Started Manual 922157-00 Rev A 9 Master Control Module Configuration This configuration includes either a 9CZi-A Master Control Module or MCM-
LabMaster 9 Zi-A 162 922157-00 Rev A OR If you selected Math On Parameters, touch inside the Math Operator data entry field and select a math function
Getting Started Manual 922157-00 Rev A 163 To select a measure mode: 1. Touch Measure → Measure Setup... on the menu bar. 2. Under Measure Mode, sel
LabMaster 9 Zi-A 164 922157-00 Rev A Standard Vertical Parameter Help Markers Turning On Markers 1. Touch Measure → Measure Setup... on the menu bar
Getting Started Manual 922157-00 Rev A 165 Turning Off Markers 1. Touch the Clear All button to turn off Help Markers for all parameters. 2. Turn of
LabMaster 9 Zi-A 166 922157-00 Rev A Analysis Teledyne LeCroy provides a variety of unique capabilities for advanced measurement parameter analysis. M
Getting Started Manual 922157-00 Rev A 167 Track and Trend are summarized in the following table: Characteristic Track Trend Representation Parameter
LabMaster 9 Zi-A 168 922157-00 Rev A In general, Track is the tool to use if you want to capture a continuous stream of data spaced closely together.
Getting Started Manual 922157-00 Rev A 169 The Select Measurement pop-up. Access the corresponding Px dialog to change other setup items as follows:
LabMaster 9 Zi-A 170 922157-00 Rev A Touch the trace descriptor label for the math trace just created for the results, and then touch the Histogram t
Getting Started Manual 922157-00 Rev A 171 SINGLE PARAMETER HISTOGRAM SETUP USING MATH Access the main Math dialog by selecting Math → Math Setup... o
LabMaster 9 Zi-A 10 922157-00 Rev A Master Acquisition Module Configuration This configuration includes a 9xxMZi-A Master Acquisition Module with corr
LabMaster 9 Zi-A 172 922157-00 Rev A Regardless of how you performed the previously described initial steps, at this point access your corresponding F
Getting Started Manual 922157-00 Rev A 173 Viewing Thumbnail Histograms (Histicons) Histicons are miniature histograms of parameter measurements that
LabMaster 9 Zi-A 174 922157-00 Rev A 5. Touch inside the Slice Center field and enter a value using the pop-up keypad. 6. Touch inside the Slice Wid
Getting Started Manual 922157-00 Rev A 175 6. Touch the newly displayed Trend math function trace label to change settings on the Trend dialog (locat
LabMaster 9 Zi-A 176 922157-00 Rev A 6. Touch the newly displayed Track math function trace label if you want to change any settings in the Track dia
Getting Started Manual 922157-00 Rev A 177 Testing and Enabling Pass/Fail Conditions Beside each Qx button on the main Pass/Fail dialog is a correspon
LabMaster 9 Zi-A 178 922157-00 Rev A When you have each of the Qx dialogs configured and enabled as desired, you can then begin your testing and turn
Getting Started Manual 922157-00 Rev A 179 Select the desired condition. NOTE: Right hand dialogs back on the corresponding Qx dialog vary based on t
LabMaster 9 Zi-A 180 922157-00 Rev A Comparing a Single Parameter 1. With the ParamCompare condition selected on a Qx dialog, touch inside the Compar
Getting Started Manual 922157-00 Rev A 181 1. Touch Analysis → Pass/Fail Setup... from the menu bar. 2. Touch one of the Qx tabs and a setup dialo
Getting Started Manual 922157-00 Rev A 11 Front of MCM-Zi Master Control Module The MCM-Zi Master Control Module includes Controls, Display, CPU, and
LabMaster 9 Zi-A 182 922157-00 Rev A The Actions Dialog Various configurations can be made to all or some of your Pass/Fail conditions using the Actio
Getting Started Manual 922157-00 Rev A 183 Depending on your oscilloscope model, you can configure up to 8 pass/fail conditions. The Boolean condition
LabMaster 9 Zi-A 184 922157-00 Rev A have the results printed (and your oscilloscope is not equipped with a printer) make sure your instrument is conn
Getting Started Manual 922157-00 Rev A 185 Math Math Traces and Functions Overview Teledyne LeCroy offers a deep toolset of math functions always grow
LabMaster 9 Zi-A 186 922157-00 Rev A You can even chain traces and/or functions to another trace or function. For instance: making Trace F1 an aver
Getting Started Manual 922157-00 Rev A 187 Regardless of how you performed the previously described initial steps, at this point access your correspon
LabMaster 9 Zi-A 188 922157-00 Rev A Mask Testing Users may load or create a pixel-based mask on the display grid using a variety of methods. Test con
Getting Started Manual 922157-00 Rev A 189 2. With the Mask test Pass Fail Condition selected, the Test, Load Mask, Make Mask, and Gate right-hand di
LabMaster 9 Zi-A 190 922157-00 Rev A Removing a Mask from the Display Quickly remove a mask by accessing the Qx dialog. On the Load Mask right-hand di
Getting Started Manual 922157-00 Rev A 191 Right-Hand Dialogs On the Test right-hand dialog, make a selection in the Test is True when group of but
LabMaster 9 Zi-A 12 922157-00 Rev A Back of MCM-Zi Master Control Module Number and Description 1. ChannelSync Outputs - SMA 10 GHz Clock and PCIe 1
LabMaster 9 Zi-A 192 922157-00 Rev A WaveScan Overview WaveScan enables you to search for unusual events in a single capture, or to scan for an event
Getting Started Manual 922157-00 Rev A 193 Signal Views WaveScan provides distinct views of your signal: Source View highlights all occurrences of
LabMaster 9 Zi-A 194 922157-00 Rev A Measurement Mode – Searches from 2 to 64 bits; ideal for bursted patterns where a PLL cannot lock. Bus Pa
Getting Started Manual 922157-00 Rev A 195 Customization Overview The instrument provides powerful capability to add your own parameters, functions, d
LabMaster 9 Zi-A 196 922157-00 Rev A Documenting Your Work Using LabNotebook Teledyne LeCroy's LabNotebook feature extends the documentation capa
Getting Started Manual 922157-00 Rev A 197 LabNotebook When LabNotebook is accessed by File → LabNotebook on the menu bar, the main LabNotebook dialog
LabMaster 9 Zi-A 198 922157-00 Rev A PLEASE NOTE THE FOLLOWING: When an entry is first created and Report Title and Description information is pro
Getting Started Manual 922157-00 Rev A 199 You can use the Up and Down arrows to move through your My Notebook Entries list one row at a time. If you
LabMaster 9 Zi-A 200 922157-00 Rev A The E-Mail button automatically sends the report data to the account you've specified in Preferences. See
Getting Started Manual 922157-00 Rev A 201 Save/Recall The Save/Recall section allows for storage and retrieval of Waveforms, Table Data, and Instrume
Getting Started Manual 922157-00 Rev A 13 The PCIe 4 Lane Data Inputs (item 8) also accommodate PCIe Expansion Slot options for GPIB and LSIB. Opti
LabMaster 9 Zi-A 202 922157-00 Rev A Saving and Recalling Setups The Save/Recall Setup dialogs allow for quick saving and recalling of up to six oscil
Getting Started Manual 922157-00 Rev A 203 RECALLING SETUP(S) DIRECTLY FROM A FILE Recall your setup(s) directly from a file by using controls on the
LabMaster 9 Zi-A 204 922157-00 Rev A 3. Destination - Click to show the available memory traces when saving to memory. 4. Touch inside the Trace Tit
Getting Started Manual 922157-00 Rev A 205 6. Depending on your selection, you may need to touch the SubFormat control and select a subformat. Byt
LabMaster 9 Zi-A 206 922157-00 Rev A Recalling Waveforms NOTE: Only .trc files saved in binary format can be recalled into the oscilloscope. Access t
Getting Started Manual 922157-00 Rev A 207 7. Up /Down arrows - Use these buttons to cycles through files in the selected folder from the Show only f
LabMaster 9 Zi-A 208 922157-00 Rev A Maximize the application by touching the shortcut icon located on the lower-right of the screen. A sequence of c
Getting Started Manual 922157-00 Rev A 209 Status Control values are shown for information about your instrument. Among them include your Hardware and
LabMaster 9 Zi-A 210 922157-00 Rev A PLEASE NOTE THE FOLLOWING: GPIB is an option that either requires a GPIB card to be installed in an open card
Getting Started Manual 922157-00 Rev A 211 Configuring the Remote Control Assistant Event Log 1. Under Remote Control Assistant, touch inside the Log
LabMaster 9 Zi-A 14 922157-00 Rev A Back of 9CZi-A Master Control Module Number and Description 1. Channel Sync Outputs - SMA 10 GHz Clock and PCIe
LabMaster 9 Zi-A 212 922157-00 Rev A 4. Touch the icon for the layout Orientation you want: portrait or landscape. 5. Touch the Hardcopy Area contro
Getting Started Manual 922157-00 Rev A 213 Printing a Screen Image You can print in one of three ways: Press the PRINT front panel button. Then, t
LabMaster 9 Zi-A 214 922157-00 Rev A Clipboard The Clipboard selection on the Hardcopy tab saves the screen image on the clipboard so you can paste a
Getting Started Manual 922157-00 Rev A 215 File The File selection on the Hardcopy tab saves the screen image as a file to storage media such as a USB
LabMaster 9 Zi-A 216 922157-00 Rev A E-mail The E-mail selection on the Hardcopy tab gives you the option to e-mail your screen images, using either t
Getting Started Manual 922157-00 Rev A 217 Aux Output The Aux Output dialog allows you to specify Auxiliary and Calibration Output details. 1. If yo
LabMaster 9 Zi-A 218 922157-00 Rev A Auxiliary Output Signals A calibration signal and the following additional signals can be output through the conn
Getting Started Manual 922157-00 Rev A 219 Calibration Output Signals A calibration signal can be output from the Probe Calibration Hook on the front
LabMaster 9 Zi-A 220 922157-00 Rev A 3. You can also set the Date and Time using Windows® by touching the Windows Date/Time button. Windows shows a D
Getting Started Manual 922157-00 Rev A 221 1. Delete a single file by first touching the Delete button on the dialog (just to make sure it's se
Getting Started Manual 922157-00 Rev A 15 PLEASE NOTE THE FOLLOWING: Cap-off unused ChannelSync SMA sockets (item 1) using the provided chain-link
LabMaster 9 Zi-A 222 922157-00 Rev A Preferences A variety of customizable preferences can be set on your instrument. Access Preferences by touching U
Getting Started Manual 922157-00 Rev A 223 Acquisition Access the Acquisition preference settings in the following manner: 1. On the Offset Setting
LabMaster 9 Zi-A 224 922157-00 Rev A This preference clears the trigger counter each time the oscilloscope issues an arm acquisition command and appli
Getting Started Manual 922157-00 Rev A 225 E-Mail Before you can send e-mail from the oscilloscope, configurations must be made on this dialog. 1. O
LabMaster 9 Zi-A 226 922157-00 Rev A Color You can customize the Colors used for Channels, Math, and Memory traces both when shown on your instrument
Getting Started Manual 922157-00 Rev A 227 Miscellaneous Other Preference setting controls are shown on the Miscellaneous dialog as follows: Numbered
LabMaster 9 Zi-A 228 922157-00 Rev A System Recovery Tool Your oscilloscope is designed to operate reliably for many years. The application software o
Getting Started Manual 922157-00 Rev A 229 Since third-party recovery software is used, our instructions certainly may not be the most definitive or c
LabMaster 9 Zi-A 230 922157-00 Rev A 5. Select Recovery from the Main Menu. The Restore Data Wizard opens. 6. Click Next on the Welcome page. 7. On
Getting Started Manual 922157-00 Rev A 231 16. Once the restoration is complete, an Information window is displayed indicating that the Data was succ
LabMaster 9 Zi-A 16 922157-00 Rev A Front of 9xxMZi-A Master Acquisition Module The 9xxMZi-A Master Acquisition Module includes Controls, Display, Ch
LabMaster 9 Zi-A 232 922157-00 Rev A Restarting the Application Upon initial power-up, the oscilloscope automatically loads the instrument application
Getting Started Manual 922157-00 Rev A 233 Reference Specifications NOTE: Specifications are subject to change without notice. Please refer to the Tel
LabMaster 9 Zi-A 234 922157-00 Rev A EN 61000-4-6:2007 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz - 80 MHz 4 EN 61000-4-11:2004 Mains Dips a
Getting Started Manual 922157-00 Rev A 235 Safety Compliance EC DECLARATION OF CONFORMITY – LOW VOLTAGE The oscilloscope meets intent of EC Directive
LabMaster 9 Zi-A 236 922157-00 Rev A CANADIAN CERTIFICATION The oscilloscope has been certified by Underwriters Laboratories (UL) to conform to the f
Getting Started Manual 922157-00 Rev A 237 Contact Teledyne LeCroy Teledyne LeCroy Service Centers United States and Canada - World Wide Corporate Of
LabMaster 9 Zi-A 238 922157-00 Rev A X-Stream® Software End-User License Agreement IMPORTANT-READ CAREFULLY: THIS END-USER LICENSE AGREEMENT ("EU
Getting Started Manual 922157-00 Rev A 239 solely by you) that are designed to operate in conjunction with an Instrument (“Your Software”), and (iii)
LabMaster 9 Zi-A 240 922157-00 Rev A the Software Product is an upgrade of a component of a package of software programs that you licensed as a single
Getting Started Manual 922157-00 Rev A 241 are necessary to achieve the Permitted Objective; (C) the information to be gained thereby has not already
Getting Started Manual 922157-00 Rev A 17 Back of 9xxMZi-A Master Acquisition Module Number and Description 1. Channel Sync Outputs - SMA 10 GHz Cl
LabMaster 9 Zi-A 242 922157-00 Rev A other supplemental software code provided to you pursuant to the Support Services will be considered part of the
Getting Started Manual 922157-00 Rev A 243 party shall either return to the disclosing party originals and copies of any Confidential Information and
LabMaster 9 Zi-A 244 922157-00 Rev A without a breach of any separate nondisclosure obligation to the disclosing party, (C) publicly known without bre
Getting Started Manual 922157-00 Rev A 245 product of the Software Product (the foregoing collectively referred to as the "Restricted Components&
LabMaster 9 Zi-A 246 922157-00 Rev A USAGE OF TRADE. YOU ACKNOWLEDGE THAT NO WARRANTIES HAVE BEEN MADE TO YOU BY OR ON BEHALF OF TELEDYNE LECROY OR OT
Getting Started Manual 922157-00 Rev A 247 obligations hereunder, of any jurisdiction in or from which you directly or indirectly cause the Software P
LabMaster 9 Zi-A 248 922157-00 Rev A 8.6. Notices. All notices or other communications between Teledyne LeCroy and you under this EULA will be in writ
Getting Started Manual Index 2.4 mm Interface, 66 2.92 mm Interface, 66 9CZi-A Master Control Module back, 14 front, 13 9xxMZi-A Master Acquisition
LabMaster 9 Zi-A 250 922157-00 Rev A disk utilities, 220 Displaying an Individual Segment, 107 Displays Containing Masks, 147 Dual Display Grid Select
Getting Started Manual 922157-00 Rev A 251 measure modes, 162 standard vertical, 162 Pass/Fail Condition Setup, 178 Pass/Fail Parameter Testing Over
LabMaster 9 Zi-A 18 922157-00 Rev A PLEASE NOTE THE FOLLOWING: Cap off unused ChannelSync SMA sockets (item 1) using the provided chain-linked 50
LabMaster 9 Zi-A 252 922157-00 Rev A displaying, 107 Send Test Mail, 225 Sequence Display Modes, 103 sequence mode display setup, 147 Sequence Samplin
Getting Started Manual 922157-00 Rev A 253 autocalibration, 222 auxiliary output signals, 218 clipboard, 214 delay control, 223 delay in, 223 email,
LabMaster 9 Zi-A 254 922157-00 Rev A
Getting Started Manual 922157-00 Rev A 19 Back of CPU Module Number and Description 1. AC Power Plug 2. Removable Hard Drive 3. The I/O Panel 4.
LabMaster 9 Zi-A 20 922157-00 Rev A Input/Output Panel 9CZi-A Master Control Module and 9xxMZi-AMaster Acquisition Module The available connections a
Getting Started Manual 922157-00 Rev A 21 MCM-Zi Master Control Module The I/O panel is located on the back of the MCM-Zi Master Control Module (page
LabMaster 9 Zi-A Oscilloscopes Getting Started Manual May, 2013
LabMaster 9 Zi-A 22 922157-00 Rev A Front of 9xxSZi-A Slave Acquisition Module Number and Description 1. ProLink Compatible Inputs 2. 2.4/2.92 mm
Getting Started Manual 922157-00 Rev A 23 Back of 9xxSZi-A Slave Acquisition Module Number and Description 1. ChannelSync PCIe 1 Lane Control Input
LabMaster 9 Zi-A 24 922157-00 Rev A ChannelSync Mainframe Hub The CMH-20 ChannelSync Mainframe Hub provides a simple and effective means to expand a L
Getting Started Manual 922157-00 Rev A 25 LabMaster Hardware Set Up Overview These instructions explain the setup of your LabMaster modules for both
LabMaster 9 Zi-A 26 922157-00 Rev A LabMaster Master Control Module and Master Acquisition Module stack setups LabMaster set up inside the availab
Getting Started Manual 922157-00 Rev A 27 9CZi-A Master Control Module Configuration This LabMaster configuration involves making PCIe 1 and 4 Lane, D
LabMaster 9 Zi-A 28 922157-00 Rev A Now, on the back of the Slave Acquisition Module, connect the other end of the same PCIe 1 Lane cable into the cor
Getting Started Manual 922157-00 Rev A 29 PLEASE NOTE THE FOLLOWING: PCIe 1 Lane Cable plugs are keyed with a single groove along one wide side of
LabMaster 9 Zi-A 30 922157-00 Rev A On the back of a Slave Acquisition Module, connect one end of a PCIe 4 Lane cable from the PCIe 4 Lane Data Output
Getting Started Manual 922157-00 Rev A 31 Gently pull the green tab on the cable plug to remove it. Repeat the previous steps for every additional Sl
© 2013 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales
LabMaster 9 Zi-A 32 922157-00 Rev A DVI-D Connection - Video Card to Touch-Screen Display on the Master Control Module The DVI-D Video Output must be
Getting Started Manual 922157-00 Rev A 33 NOTE: The connections are specifically DVI-D. SMA 72" Cables - Master Control Module to Slave Acquisit
LabMaster 9 Zi-A 34 922157-00 Rev A On the back of the Master Control Module, connect one end of an SMA 72" cable from a SMA 10 GHz Clock Output
Getting Started Manual 922157-00 Rev A 35 Cap off any unused SMA 10 GHz Clock Outputs on the back of your 9CZi-A Master Control Module using the chain
LabMaster 9 Zi-A 36 922157-00 Rev A When all connections are made, the main switch on the front of the Master Control Module powers all connected it
Getting Started Manual 922157-00 Rev A 37
LabMaster 9 Zi-A 38 922157-00 Rev A MCM-Zi Master Control Module Configuration This LabMaster configuration involves making PCIe 1 and 4 Lane, SMA 72&
Getting Started Manual 922157-00 Rev A 39 Now, connect the other end of the 1 Lane cable into the SYNC socket on the back of the Acquisition Module.
LabMaster 9 Zi-A 40 922157-00 Rev A PLEASE NOTE THE FOLLOWING: PCIe 1 Lane Cable plugs are keyed with a single groove along one wide side of the p
Getting Started Manual 922157-00 Rev A 41 On the back of the Acquisition Module, connect one end of a PCIe 4 lane cable from the PCIe 4 Lane Data Outp
Getting Started Manual 922157-00 Rev A i TABLE OF CONTENTS Welcome ...
LabMaster 9 Zi-A 42 922157-00 Rev A Gently pull the green tab on the cable plug to remove it. Repeat the previous steps for every additional Acquisit
Getting Started Manual 922157-00 Rev A 43 SMA 72" Cables - ChannelSync Clock Connections Connect each SMA 10 GHz Clock Output on the back of your
LabMaster 9 Zi-A 44 922157-00 Rev A Now, on the back of the Acquisition Module, connect the other end of the same SMA 72" cable into the correspo
Getting Started Manual 922157-00 Rev A 45 Power Cable Connections and Main Power Switch Teledyne LeCroy advises that you note the power ratings of the
LabMaster 9 Zi-A 46 922157-00 Rev A 9xxMZi-A Master Acquisition Module Configuration PCIe 1 Lane Cable - Master Acquisition Module to Slave Acquisitio
Getting Started Manual 922157-00 Rev A 47 Gently pull the green tab on the cable plug to remove it. Repeat the previous steps for every additional S
LabMaster 9 Zi-A 48 922157-00 Rev A PCIe 4 Lane Cable - CPU Module to Master Acquisition Module Make a connection for Channels 1-4 from the back of th
Getting Started Manual 922157-00 Rev A 49 Now, on the back of the CPU Module, connect the other end of the same PCIe 4 Lane cable specifically into th
LabMaster 9 Zi-A 50 922157-00 Rev A PLEASE NOTE THE FOLLOWING: PCIe 4 Lane Cable plugs are keyed with a single groove along one wide side of the p
Getting Started Manual 922157-00 Rev A 51 Plug one end of a PCIe 4 Lane cable into the PCIe 4 Lane Data Output (labeled DATALINK) on the back of each
LabMaster 9 Zi-A ii 922157-00 Rev A Vertical Controls ... 7
LabMaster 9 Zi-A 52 922157-00 Rev A Gently pull the green tab on the cable plug to remove it. PLEASE NOTE THE FOLLOWING: PCIe 4 Lane Cable plugs
Getting Started Manual 922157-00 Rev A 53 DVI-D Connection - CPU Module's Video Card Output to the Master Acquisition Module's Touch-Screen
LabMaster 9 Zi-A 54 922157-00 Rev A Now, connect the other end of the same DVI-D cable into the DVI-D Video Input on the back of the Master Acquisitio
Getting Started Manual 922157-00 Rev A 55 SMA 72" Cables - Master Acquisition Module to Slave Acquisition Module Reference Clock Connection Conne
LabMaster 9 Zi-A 56 922157-00 Rev A Now, on the back of the Slave Acquisition Module, connect the other end of the same SMA 72" cable into the co
Getting Started Manual 922157-00 Rev A 57 Power Cable Connection and Main Power Switch The combined draw from your 9xxMZi-AMaster Acquisition Module,
LabMaster 9 Zi-A 58 922157-00 Rev A When all connections are made, the Standby Power switch on the front of the Master Acquisition Module powers all
Getting Started Manual 922157-00 Rev A 59 Removing and Attaching the Front Panel Control Detach the Front Panel Control from the oscilloscope by slid
LabMaster 9 Zi-A 60 922157-00 Rev A FRONT PANEL AS A REMOTE CONTROL While detached, your front panel (standard or 4 channel version) can be used as a
Getting Started Manual 922157-00 Rev A 61 Touch Screen and External Displays Set Up Internal Display Connect the DVI-D Video Output on the CPU to the
Getting Started Manual 922157-00 Rev A iii Viewing Waveforms ...
LabMaster 9 Zi-A 62 922157-00 Rev A 5. Use the Start DSO desktop icon to restart the oscilloscope application. 6. Choose Display > Display Setup.
Getting Started Manual 922157-00 Rev A 63 Grids distributed after Dual Display Mode selection. Trace descriptor boxes move with the trace (note the m
LabMaster 9 Zi-A 64 922157-00 Rev A Signal Inputs Interfaces LabMaster 9 Zi systems are equipped with a variety of interfaces to allow you to connect
Getting Started Manual 922157-00 Rev A 65 control for a wide range of probes with bandwidth ratings from 3 GHz to 20 GHz. The nature of the ProLink in
LabMaster 9 Zi-A 66 922157-00 Rev A combination of grooves, external threads and a coaxial interface with a 2.92 mm airline geometry. The second c
Getting Started Manual 922157-00 Rev A 67 NOTE: The mounted connectors for the 2.92 mm and 2.4 mm interfaces look identical, and the connector savers
LabMaster 9 Zi-A 68 922157-00 Rev A First, the BMA Female Connector (1). Then, shown installed are the ProLink LPA-SMA, LPA-BNC, and LPA-K Adapters
Getting Started Manual 922157-00 Rev A 69 designed for connection to a 2.92 mm Probe Connection Interface to be connected to a ProLink input connectio
LabMaster 9 Zi-A 70 922157-00 Rev A After a probe is connected, it is recognized and an additional tab with the probe model name is displayed to the r
Getting Started Manual 922157-00 Rev A 71 Passive Probe Compensation Passive probes must be compensated to flatten overshoot. This is accomplished by
LabMaster 9 Zi-A iv 922157-00 Rev A Parameter Measurements ... 194 Sampling
LabMaster 9 Zi-A 72 922157-00 Rev A Front Panel The following Basic Control topics cover the general usage of the hardware buttons located on the osci
Getting Started Manual 922157-00 Rev A 73 Miscellaneous Controls and WaveStream Indicator Help - Press to open the Teledyne LeCroy Online Assistant
LabMaster 9 Zi-A 74 922157-00 Rev A Press the AUTO SETUP... front panel button twice to perform the last selection from the Auto Setup... flyout menu
Getting Started Manual 922157-00 Rev A 75 Auto - Press to turn on Auto Trigger mode, which triggers the oscilloscope after a time-out, even if the
LabMaster 9 Zi-A 76 922157-00 Rev A Vertical Controls NOTE: You can turn channels on and off using the software as explained in Vertical Overview (on
Getting Started Manual 922157-00 Rev A 77 Zoom and Math Controls NOTE: Zoom and Math front panel controls correspond with screen menu selection: Math
LabMaster 9 Zi-A 78 922157-00 Rev A Display Dashboard This section covers the various aspects of the user interface shown in the display of your instr
Getting Started Manual 922157-00 Rev A 79 Menu Bar The top of the screen contains a menu bar of commonly used functions. Whenever you touch one of the
LabMaster 9 Zi-A 80 922157-00 Rev A Stop - Press to prevent the oscilloscope from triggering on a signal. If you boot up the instrument with the tr
Getting Started Manual 922157-00 Rev A 81 Signal Display Grid The grid area is divided into 8 vertical divisions and 10 horizontal divisions just like
Getting Started Manual 922157-00 Rev A 1 Welcome Thank you for purchasing a Teledyne LeCroy product. We're certain you'll be pleased with th
LabMaster 9 Zi-A 82 922157-00 Rev A The default setting (Time) is for delay readout (in seconds) and to move proportionately when the timebase knob is
Getting Started Manual 922157-00 Rev A 83 Signal Display Grid Pop-Up Menu On the Signal Display Grid, the Pop-up menu provides assistance while using
LabMaster 9 Zi-A 84 922157-00 Rev A The long and short forms of trace descriptor indicators. Besides channel traces, math and parameter measurement l
Getting Started Manual 922157-00 Rev A 85 Shortcut Buttons You can access the same functions as the ones produced when using the Signal Display Grid P
LabMaster 9 Zi-A 86 922157-00 Rev A A dialog box opens in which to create the label. The first time creating a waveform label, Label1 is provided as d
Getting Started Manual 922157-00 Rev A 87 O.K. on the keyboard when you are done. The edited text automatically appears in the label on the waveform.
LabMaster 9 Zi-A 88 922157-00 Rev A Other times, a larger box is shown after touching a control. This larger Pop-Up has categorical buttons along the
Getting Started Manual 922157-00 Rev A 89 FOLDER/FILE BROWSING CONTROLS These controls allow for navigation to or from folders (on the hard drive or m
LabMaster 9 Zi-A 90 922157-00 Rev A FLYOUT MENU CONTROLS Flyout Menus provides a variety of solutions for a particular main area of functionality. It
Getting Started Manual 922157-00 Rev A 91 Text entry field for LabNotebook using the Virtual Keyboard. Slider Bar Some models provide what is known a